Mission: To establish and facilitate partnerships that provide a framework for disaster readiness and continuity of care for the renal community. The Florida Kidney Disaster Coalition is the model disaster planning and readiness organization for kidney patients by collaborating with community partners to ensure timely access to quality care.
Wednesday, December 1, 2010
Recall Notice: B. Braun addEASE Binary Connector: Class I Recall -Stopper Fragments May Enter Bag
On November 30, 2010, B Braun Medical Inc. issued a reiteration of a prior June 28, 2010 recall on use of its addEASE Binary Connectors. The organization identified that when an addEASE Binary Connector is inserted into a partial additive bag (PAB) stopper, fragments of the stopper may enter the bag, resulting in a small amount of visible particles in the solution. These particles can potentially enter a patient’s body and lead to serious injury and/or death. It is recommended that anyone currently using this product discontinue doing so immediately. Click here for more information.